FDA Enforcement Class II Ongoing

Sensica Urine Output System, Catalog Number SCCS1002

Recall: Z-0554-2024 · Reported December 20, 2023

Enforcement

Recall Number
Z-0554-2024
Event ID
93375
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
C.R. Bard Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 20, 2023
Initiation Date
November 15, 2023
Classification Date
December 14, 2023
Address
8195 Industrial Blvd Ne, N/A, Covington, GA, 30014-1497, United States

Description

Sensica Urine Output System, Catalog Number SCCS1002

Reason

BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.

Code Info

UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007, BMGPRMYS020, BMGPRMYS026, BMGPRMYS028, BMGPRMYS030, BMFWRMSBT4, BMFWRMSJ45, BMFWRMSJ49, BMFZRMS104, BMGNRMS085, BMFSRMSH98, BMFSRMSJ01, BMFVRMSQ62, BMFVRMSQ66, BMFVRMSQ68, BMFYRMSJ41, BMFYRMSJ49, BMFZRMS067, BMFZRMS070, BMFZRMS080, BMFZRMS113, BMFZRMS117, BMGNRMS070, BMFURMSF02, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ89, BMFVRMSQ91, BMFWRSJ46, BMGPRMYS011, BMGPRMYS018, BMGPRMYS022 , BMGPRMYS023, BMGPRMYS027, BMFYRMSJ35, BMFYRMSJ37, BMFYRMSJ54, BMFYRMSJ58, BMGNRMS079, BMGNRMSG69, BMGNRMSG74, BMFSRMSH94, BMFSRMSH99, BMFSRMSJ38, BMFSRMSJ44, BMFTRMS138, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFQRMS248, BMFQRMS250, BMFRRMS646, BMFRRMS647, BMGPRMYS008, BMGPRMYS009, BMGPRMYS010, BMGPRMYS013, BMGPRMYS014, BMGPRMYS016, BMFZRMS073, BMFSRMSH78, BMFSRMSH79, BMFSRMSH80, BMFSRMSH81, BMFSRMSH82, BMFSRMSH83, BMFSRMSH93, BMFSRMSJ00, BMFSRMSJ14, BMFSRMSJ15, BMFQRMS246, BMFWRMSJ39, BMFWRMSJ41, BMFSRMSJ36, BMFSRMSJ37, BMFSRMSJ42, BMFTRMS106, BMFTRMS109, BMFTRMS114, BMFTRMS142, BMFURMSF01, BMFURMSF05, BMFSRMSH95, BMFSRMSH97, BMFTRMS120, BMFTRMS122, BMFTRMS123, BMFTRMS135, BMFTRMS136, BMFTRMS137, BMGPRMYS002, BMGPRMYS012, BMGPRMYS019 , BMGPRMYS029, BMFPRMYS009, BMFQRMS244, BMFQRMS272, BMFQRMS276, BMFQRMS286, BMFQRMS293, BMFSRMSJ43, BMFTRMS121, BMGNRMS067, BMGNRMS076, BMFVRMSQ58, BMFVRMSQ70, BMFVRMSQ73, BMFVRMSQ74, BMFVRMSQ90, BMFVRMSR17, BMFWRMSJ27, BMFWRMSJ35, BMFURMSF06, BMFURMSF09, BMFWRMSJ36, BMFWRMSJ51, BMFZRMS068, BMFQRMS294, BMFZRMS105, BMFSRMSJ20, BMFTRMS140, BMFVRMSQ63, BMFVRMSQ64, BMFVRMSQ65, BMFVRMSQ72, BMFVRMSQ75, BMFVRMSQ76, BMFVRMSQ77, BMFVRMSQ78, BMFVRMSQ80, BMFYRMSJ39, BMFZRMS081, BMFZRMS116, BMGNRMS015, BMGNRMS029, BMGNRMS073

Distribution

US Nationwide distribution.

Quantity

148 units