Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.
Enforcement
- Recall Number
- Z-0554-2021
- Event ID
- 86846
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Shimadzu Medical Systems
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 16, 2020
- Initiation Date
- November 18, 2020
- Classification Date
- December 10, 2020
- Address
- 20101 S Vermont Ave, N/A, Torrance, CA, 90502-1328, United States
Description
Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.
The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.
Serial Numbers: 40AD58C55001, 40AD58C4B001, 40AD58C24001, 40AD58522001, 40AD58C1C001, 101X968101, 091X955901
US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.
7 units