FDA Enforcement Class II Terminated

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

Recall: Z-0554-2018 · Reported February 14, 2018

Enforcement

Recall Number
Z-0554-2018
Event ID
79005
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mindray DS USA, Inc. dba Mindray North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 14, 2018
Initiation Date
July 14, 2017
Classification Date
February 8, 2018
Termination Date
December 17, 2018
Address
800 Macarthur Blvd, N/A, Mahwah, NJ, 07430-2001, United States

Description

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

Reason

A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.

Code Info

All battery chargers with part number 115-030108-00

Distribution

Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.

Quantity

203