FDA Enforcement Class II Terminated

ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

Recall: Z-0553-2022 · Reported February 9, 2022

Enforcement

Recall Number
Z-0553-2022
Event ID
89244
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Luminex Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 9, 2022
Initiation Date
November 17, 2021
Classification Date
February 2, 2022
Termination Date
July 8, 2024
Address
12212 Technology Blvd, N/A, Austin, TX, 78727-6101, United States

Description

ARIES HSV 1&2 Assay, Catalog/REF# 50-10017

Reason

Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.

Code Info

Lot Numbers AB3600A and AB3560A

Distribution

Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.

Quantity

329