FDA Enforcement
Class II
Ongoing
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.
Recall: Z-0542-2026
·
Reported November 26, 2025
Enforcement
- Recall Number
- Z-0542-2026
- Event ID
- 97835
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Cellavision AB
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 26, 2025
- Initiation Date
- October 8, 2025
- Classification Date
- November 18, 2025
- Address
- Forskningsbyn Ideon, Scheelevagen 19a, Lund, N/A, Sweden
Description
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.
Reason
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Code Info
UDI: 0735004097202/ Serial Numbers 63220-70093 or CN63220-CN70093 Serial Numbers 60001-63219 or CN60001-CN63219
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.
Quantity
688 units