FDA Enforcement Class II Terminated

Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument back and forth until the entire tab is fully detached from the screw head. Catalog number: 5101-90195

Recall: Z-0542-2021 · Reported December 16, 2020

Enforcement

Recall Number
Z-0542-2021
Event ID
86733
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
K2M, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 16, 2020
Initiation Date
November 4, 2020
Classification Date
December 7, 2020
Termination Date
February 22, 2024
Address
600 Hope Pkwy SE, N/A, Leesburg, VA, 20175-4428, United States

Description

Everest MI XT Tab Removal Tool (i.e. All-In-One Tab Removal Tool)-used to break the tab off the Everest XT screw by rocking the instrument back and forth until the entire tab is fully detached from the screw head. Catalog number: 5101-90195

Reason

All-in-One Tab Removal Tool may cause cracking and/or separating at the threaded interface between tool s metal shaft component and handle, resulting in leakage of Loctite from handle cracks

Code Info

Lot Numbers: JUFT, JUGX, KEFE, KUBJ, KVNT, KYRR, MAJR GTIN: 10888857290679

Distribution

Nationwide Foreign: Netherlands, Spain, UK, Belgium, Germany, Italy, Norway and Sweden

Quantity

197 units