FDA Enforcement Class III Terminated

THERMO FISHER 8 DRUG CASSETTE, Item No. MTPA-8MBAU

Recall: Z-0542-2016 · Reported December 30, 2015

Enforcement

Recall Number
Z-0542-2016
Event ID
72651
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Ameditech Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 30, 2015
Initiation Date
November 16, 2015
Classification Date
December 24, 2015
Termination Date
November 9, 2016
Address
9940 Mesa Rim Rd, N/A, San Diego, CA, 92121-2910, United States

Description

THERMO FISHER 8 DRUG CASSETTE, Item No. MTPA-8MBAU

Reason

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code Info

G150328

Distribution

Nationwide Distribution.

Quantity

100 kits