FDA Enforcement Class I Ongoing

AirTouch N20 Nasal Mask and User Guide

Recall: Z-0540-2024 · Reported December 27, 2023

Enforcement

Recall Number
Z-0540-2024
Event ID
93499
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
ResMed Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 27, 2023
Initiation Date
November 20, 2023
Classification Date
December 21, 2023
Address
1 Elizabeth Macarthur Dr, N/A, Bella Vista, N/A, N/A, Australia

Description

AirTouch N20 Nasal Mask and User Guide

Reason

Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

Code Info

All mask lots used with User Guide 638218/version 2020-02, and prior. UDI-DI/GTIN: 619498639069, 619498639045, 619498639052, 619498639021, 619498639038, 619498639007, 619498639076, 619498639014, 619498639540, 619498639205, 619498639229, 619498639113, 619498639090, 619498639120, 619498639212

Distribution

US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA

Quantity

424,309