AirFit N20 Nasal Mask and User Guide
Enforcement
- Recall Number
- Z-0539-2024
- Event ID
- 93499
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ResMed Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 27, 2023
- Initiation Date
- November 20, 2023
- Classification Date
- December 21, 2023
- Address
- 1 Elizabeth Macarthur Dr, N/A, Bella Vista, N/A, N/A, Australia
Description
AirFit N20 Nasal Mask and User Guide
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
All mask lots used with User Guide 638218/version 2020-02, 638243/version 2020-06, and prior. UDI-DI/GTIN: 619498635016, 619498635054, 619498635061, 619498635351, 619498635047, 619498635030, 619498635078, 619498635382, 619498635023, 619498635344, 619498635009, 619498635399, 619498635115, 619498635122, 619498635375, 619498635214, 619498635221, 619498635252, 619498635245, 619498635313, 619498635337, 619498635207, 619498635320, 619498635238, 619498635092, 619498635108, 619498635306, 619498635290, 619498635184, 619498635146, 619498635160, 619498635139, 619498635177, 619498635153
US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA
7,724,968