FDA Enforcement
Class III
Ongoing
Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen (equine tendon) For In-Vitro Diagnostic Use,
Recall: Z-0537-2020
·
Reported December 4, 2019
Enforcement
- Recall Number
- Z-0537-2020
- Event ID
- 83930
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Helena Laboratories, Corp.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- December 4, 2019
- Initiation Date
- May 23, 2018
- Classification Date
- November 22, 2019
- Address
- 1530 Lindbergh Dr, Beaumont, TX, 77707-4131, United States
Description
Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen (equine tendon) For In-Vitro Diagnostic Use,
Reason
Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet aggregation Kit, may contain particulate matter or microbial growth.
Code Info
Component Lot # 5-17-5368 (Collagen), Exp. 5/31/19 Kit Lot # 3-17-5369
Distribution
US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan
Quantity
64 Aggregation Kits