VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
Enforcement
- Recall Number
- Z-0535-2025
- Event ID
- 95634
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ortho-Clinical Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- December 4, 2024
- Initiation Date
- October 11, 2024
- Classification Date
- November 25, 2024
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States
Description
VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.
UDI/GTIN 10758750000296 Coating 1221, expiration date 10/01/2024 Coating 1222, expiration date 01/01/2025
Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
7717 units