CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Enforcement
- Recall Number
- Z-0532-2021
- Event ID
- 86650
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Canon Medical System, USA, INC.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- FAX
- Report Date
- December 9, 2020
- Initiation Date
- October 8, 2020
- Classification Date
- December 3, 2020
- Termination Date
- December 6, 2022
- Address
- 2441 Michelle Dr, N/A, Tustin, CA, 92780-7047, United States
Description
CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.
Affected Serial Numbers: 1 AFA1662001, 2 AFB1872005, 3 AFB1832003, 4 AFB1872004, 5 AFB1892006, 6 BFA18Y7001, 7 AEA17X2001, 8 AFA1692002, 9 AEB1812002, 10 BFA1932001, 11 BFB1952003, 12 BFB1942002, 13 BFB1972004, 14 BFB1972005, 15 AFB1892007, 16 BEC2032001, 17 BFC2062007, 18 BFC2062008, 19 BHA1992001, 20 BFC2012006
US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.
14 units