Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.
Recall: Z-0530-2022
·
Reported February 9, 2022
Enforcement
- Recall Number
- Z-0530-2022
- Event ID
- 89017
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 9, 2022
- Initiation Date
- December 29, 2021
- Classification Date
- January 28, 2022
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.
Reason
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Code Info
All serial numbers, GTIN 00085412498683.
Distribution
Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.
Quantity
146,502 units