FDA Enforcement
Class II
Terminated
BD BBL Trypticase Soy Agar with 5% Sheep Blood, Catalog No. 221261, (non-selective, non-differential)
Recall: Z-0528-2019
·
Reported December 12, 2018
Enforcement
- Recall Number
- Z-0528-2019
- Event ID
- 81572
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 12, 2018
- Initiation Date
- October 22, 2018
- Classification Date
- November 30, 2018
- Termination Date
- April 3, 2020
- Address
- 7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States
Description
BD BBL Trypticase Soy Agar with 5% Sheep Blood, Catalog No. 221261, (non-selective, non-differential)
Reason
a portion of a single lot was not manufactured in accordance with the internal formulation resulting in inhibited growth of Gram negative organisms.
Code Info
UDI - (01)10382902212611(17)181130 (10)8221508 Lot No.: 8221508 Expiration Date: 2018-11-30
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CT, FL, GA, HI, MA, MD, MI, MN, NC, NH, NJ, NY, OH, OK, PA, RI, TX, VA, VT, WA, WV
Quantity
54,300 plates