FDA Enforcement Class II Ongoing

smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP

Recall: Z-0527-2025 · Reported December 4, 2024

Enforcement

Recall Number
Z-0527-2025
Event ID
95605
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Smiths Medical ASD, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2024
Initiation Date
October 28, 2024
Classification Date
November 25, 2024
Address
6000 Nathan Ln N, Minneapolis, MN, 55442-1690, United States

Description

smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP

Reason

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

Code Info

REF 14-504-17, UDI/DI 35021239002579, Lot Numbers: 0003340, 0003620, 0003616, 0003580, 0003542, 0003509, 0003579, 0003568, 0003599, 0003615, 0003589, 0003562, 0003601, 0003618, 0003583, 0003613, 0003270, 0003629, 0003293, 0003584, 0003487, 0003617, 0003280, 0003348, 0003470, 0003523, 0003607, 0003576, 0003422, 0003614, 0003633, 0003578, 0003630, 0003207, 0003438, 0003559, 0003570, 0003478, 0003533, 0003588, 0003433, 0003556, 0003561, 0003446, 0003284, 0003545, 0003461, 0003590, 0003398, 0003530, 0003539, 0003550, 0003213, 0003598, 0003623, 0003504, 0003548, 0003552, 0003631, 0003594, 0003625, 0003600, 0003378, 0003526, 0003574, 0003411, 0003471, 0003587, 0003596, 0003297, 0003298, 0003412, 0003424, 0003425, 0003517, 0003151, 0003163, 0003235, 0003347, 0003350, 0003382, 0003390, 0003416, 0003423, 0003603, 0003638, 0003641, 0003175, 0003214, 0003238, 0003253, 0003257, 0003343, 0003351, 0003388, 0003397, 0003428, 0003486, 0003488, 0003496, 0003498, 0003528, 0003541, 0003608, 0003610, 0003612, 0003639, 0003640, 0003121, 0003122, 0003123, 0003125, 0003126, 0003127, 0003128, 0003129, 0003131, 0003132, 0003133, 0003134, 0003135, 0003136, 0003137, 0003138, 0003141, 0003142, 0003143, 0003145, 0003146, 0003147, 0003149, 0003150, 0003153, 0003154, 0003155, 0003156, 0003157, 0003158, 0003159, 0003160, 0003164, 0003166, 0003167, 0003168, 0003170, 0003171, 0003172, 0003173, 0003174, 0003176, 0003177, 0003178, 0003179, 0003180, 0003181, 0003182, 0003183, 0003188, 0003190, 0003191, 0003192, 0003193, 0003194, 0003195, 0003197, 0003198, 0003199, 0003200, 0003201, 0003202, 0003203, 0003208, 0003209, 0003211, 0003212, 0003215, 0003217, 0003219, 0003220, 0003221, 0003223, 0003224, 0003225, 0003227, 0003228, 0003229, 0003230, 0003231, 0003233, 0003234, 0003237, 0003239, 0003240, 0003241, 0003242, 0003243, 0003244, 0003245, 0003246, 0003247, 0003248, 0003249, 0003250, 0003251, 0003252, 0003258, 0003259, 0003260, 0003261, 0003262, 0003263, 0003264, 0003265, 0003266, 0003268, 0003269, 0003271, 0003272, 0003273, 0003274, 0003275, 0003278, 0003279, 0003281, 0003282, 0003283, 0003285, 0003286, 0003287, 0003288, 0003289, 0003290, 0003291, 0003292, 0003294, 0003295, 0003296, 0003299, 0003302, 0003303, 0003304, 0003305, 0003306, 0003309, 0003310, 0003311, 0003312, 0003313, 0003316, 0003317, 0003318, 0003319, 0003320, 0003321, 0003323, 0003324, 0003325, 0003326, 0003327, 0003328, 0003330, 0003331, 0003332, 0003333, 0003334, 0003335, 0003336, 0003337, 0003338, 0003341, 0003342, 0003344, 0003345, 0003346, 0003349, 0003352, 0003353, 0003354, 0003355, 0003356, 0003365, 0003366, 0003379, 0003383, 0003384, 0003385, 0003386, 0003387, 0003389, 0003392, 0003396, 0003400, 0003401, 0003402, 0003404, 0003405, 0003406, 0003407, 0003408, 0003409, 0003410, 0003413, 0003417, 0003418, 0003419, 0003420, 0003421, 0003426, 0003427, 0003429, 0003430, 0003431, 0003432, 0003434, 0003435, 0003436, 0003437, 0003439, 0003440, 0003441, 0003442, 0003443, 0003444, 0003445, 0003447, 0003448, 0003450, 0003451, 0003452, 0003453, 0003454, 0003455, 0003458, 0003460, 0003462, 0003463, 0003464, 0003465, 0003466, 0003467, 0003468, 0003472, 0003473, 0003474, 0003475, 0003479, 0003481, 0003482, 0003485, 0003490, 0003491, 0003493, 0003494, 0003495, 0003497, 0003499, 0003501, 0003502, 0003503, 0003505, 0003506, 0003507, 0003508, 0003512, 0003513, 0003514, 0003515, 0003516, 0003518, 0003519, 0003520, 0003521, 0003524, 0003525, 0003527, 0003531, 0003532, 0003534, 0003535, 0003536, 0003537, 0003538, 0003543, 0003544, 0003547, 0003551, 0003554, 0003560, 0003565, 0003566, 0003582, 0003586, 0003591, 0003592, 0003593, 0003595, 0003602, 0003604, 0003606, 0003611, 0003624, 0003632, 0003636, 0003637, 0003642, 0003575, 0003540, 0003572, 0003558, 0003555, 0003567, 0003585, 0003546, 0003609, 0003564, 0003563, 0003571, 0003577, 0003569; REF 14-504-17JP, UDI/DI 35021239002579, Lot Numbers: 0003215, 0003415, 0003483

Distribution

Worldwide distribution.

Quantity

142327 units