FDA Enforcement
Class II
Terminated
Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.
Recall: Z-0526-2015
·
Reported December 24, 2014
Enforcement
- Recall Number
- Z-0526-2015
- Event ID
- 69637
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Varian Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 24, 2014
- Initiation Date
- April 30, 2014
- Classification Date
- December 16, 2014
- Termination Date
- March 16, 2015
- Address
- 911 Hansen Way, N/A, Palo Alto, CA, 94304-1028, United States
Description
Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.
Reason
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
Code Info
Part Number P1003290001
Distribution
Worldwide Distribution.
Quantity
32 units