FDA Enforcement Class II Terminated

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Recall: Z-0526-2015 · Reported December 24, 2014

Enforcement

Recall Number
Z-0526-2015
Event ID
69637
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Varian Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 24, 2014
Initiation Date
April 30, 2014
Classification Date
December 16, 2014
Termination Date
March 16, 2015
Address
911 Hansen Way, N/A, Palo Alto, CA, 94304-1028, United States

Description

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Reason

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

Code Info

Part Number P1003290001

Distribution

Worldwide Distribution.

Quantity

32 units