Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Recall: Z-0523-2015
·
Reported December 17, 2014
Enforcement
- Recall Number
- Z-0523-2015
- Event ID
- 69686
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 17, 2014
- Initiation Date
- October 28, 2014
- Classification Date
- December 5, 2014
- Termination Date
- July 20, 2015
- Address
- 51 Valley Stream Pkwy, N/A, Malvern, PA, 19355, United States
Description
Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Reason
It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
Code Info
model number: 10140000
Distribution
US Nationwide Distribution in the state of: WV, NY, and FL..
Quantity
3