FDA Enforcement Class II Terminated

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Recall: Z-0523-2015 · Reported December 17, 2014

Enforcement

Recall Number
Z-0523-2015
Event ID
69686
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 17, 2014
Initiation Date
October 28, 2014
Classification Date
December 5, 2014
Termination Date
July 20, 2015
Address
51 Valley Stream Pkwy, N/A, Malvern, PA, 19355, United States

Description

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Reason

It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.

Code Info

model number: 10140000

Distribution

US Nationwide Distribution in the state of: WV, NY, and FL..

Quantity

3