FDA Enforcement
Class II
Terminated
Total Temporomandibular Joint Replacement System- 55mm Rt Standard Ti Mand Reconstruction of the temporomandibular joint
Recall: Z-0520-2015
·
Reported December 17, 2014
Enforcement
- Recall Number
- Z-0520-2015
- Event ID
- 69644
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet Microfixation, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 17, 2014
- Initiation Date
- October 17, 2014
- Classification Date
- December 8, 2014
- Termination Date
- January 4, 2017
- Address
- 1520 Tradeport Dr, N/A, Jacksonville, FL, 32218-2480, United States
Description
Total Temporomandibular Joint Replacement System- 55mm Rt Standard Ti Mand Reconstruction of the temporomandibular joint
Reason
Laser etching on the parts is wider and deeper than the conditions previously validated.
Code Info
Model 24-6555TI, Lot #s: 525210B, 525200A, 525200B, 525210A.
Distribution
Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.
Quantity
42