FDA Enforcement Class II Ongoing

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.

Recall: Z-0518-2024 · Reported December 20, 2023

Enforcement

Recall Number
Z-0518-2024
Event ID
93390
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Riverpoint Medical, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 20, 2023
Initiation Date
October 25, 2023
Classification Date
December 8, 2023
Address
825 NE 25th Ave, N/A, Portland, OR, 97232-2304, United States

Description

Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.

Reason

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

Code Info

Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617

Distribution

U.S. Nationwide distribution in the state of CA.

Quantity

999 devices