Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite¿ 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite¿ 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Console with Connector, Catalog No.: 9770600 Site~Rite 8 Ultrasound System Console with Connector, Refurbished, Catalog No: 9770600X Site~Rite 8 Ultrasound System Console with Connector (Remanufactured) Catalog No: 9770600Y
Enforcement
- Recall Number
- Z-0518-2020
- Event ID
- 84262
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton Dickinson & Company
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 4, 2019
- Initiation Date
- November 4, 2019
- Classification Date
- November 22, 2019
- Termination Date
- November 19, 2020
- Address
- 1 Becton Dr, Franklin Lakes, NJ, 07417-1815, United States
Description
Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8 Ultrasound System, Catalog No.9770500 Refurbished, Site~Rite¿ 8 System, Catalog No.9770500R Site~Rite 8 Ultrasound System, Catalog No.9770501 Refurbished Site~Rite¿ 8 System,Catalog No.9770501R Site~Rite 8 Ultrasound System, Catalog No. 9770550 Site~Rite 8 Ultrasound System, Catalog No. 9770552 Site~Rite 8 Ultrasound System, Catalog No. 9770553 Site~Rite 8 Ultrasound System, Catalog No. 9770554 Site~Rite 8 Ultrasound System, Catalog No. 9770555 Site~Rite 8 Ultrasound System Console with Connector, Catalog No.: 9770600 Site~Rite 8 Ultrasound System Console with Connector, Refurbished, Catalog No: 9770600X Site~Rite 8 Ultrasound System Console with Connector (Remanufactured) Catalog No: 9770600Y
The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only
All Serial numbers
Nationwide Foreign: UK, Belgium, Australia, Japan, Canada, Taiwan, Colombia, Chile, Philippines, Thailand, Brazil, Hong Kong, Chile, Argentina, Mexico, India, Singapore
2215 units