FDA Enforcement Class II Terminated

UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

Recall: Z-0516-2020 · Reported December 4, 2019

Enforcement

Recall Number
Z-0516-2020
Event ID
84186
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 4, 2019
Initiation Date
October 7, 2019
Classification Date
November 22, 2019
Termination Date
April 27, 2021
Address
11800 SW 147th Ave, Miami, FL, 33196-2500, United States

Description

UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

Reason

When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.

Code Info

Catalog Number 775222. Serial numbers: AS39017, AS40018, AT35013, AT35014, AT35015, AT35017, AT35018, AT40020, AT40021, AT40022, AT40024, AT40025, AT40027, AT40030, AT45031, AT45032, AT45034, AT45035, AT49037, AT49041, AT49044, AT49045, AT49046, AT49047, AT49048, AT49049, AT49050, AT49051, AT49055, AT49057, AT49058, AT49059, AT49060, AT49061, AT49062, AT49063, AT49064, AT49065, AU01004, AU01005, AU01007, AU01009, AU01010, AU01011, AU01012, AU01014, AU01015, AU01016, AU01017, AU01020, AU01021, AU01022, AU01023, AU01024, AU01025, AU01028, AU01029, AU01030, AU08041, AU08043, AU08044, AU08047, AU08048, AU08049, AU08050, AU11051, AU11053, AU11057, AU11058, AU11059, AU11061, AU11062, AU11063, AU11064, AU11066, AU11068, AU11072, AU13073, AU13075, AU13077, AU13079, AU14083, AU14084, AU14085, AU14086, AU14087, AU14088, AU14090, AU16104, AU16111, AU16112, AU19113, AU19115, AU19118, AU19119, AU19120, AU20123, AU20124, AU20125, AU20128, AU20129, AU20130, AU20131, AU21133, AU21136, AU21137, AU21141, AU23146, AU23147, AU23149, AU23150, AU23151, AU23152, AU28153, AU28154, AU28160, AU28161, AU29163, AU29165, AU29167, AU29168, AU29169, AU29172, AU30173, AU30178, AU30180, AU30181, AU30182, AU35183, AU35185, AU35186, AU35187, AU35188, AU36189, AU36192, AU36193, AU36196, AU36198, AU36200, AU36201, AU36202, AU36203, AU36204, AU40205, AU40207, AU40209, AU40210, AU40211, AU40213, AU40214, AU40215, AU40216, AU40217, AU40218, AU40221, AU43223, AU43225, AU43226, AU43228, AU43229, AU43231, AU43232, AU43233, AU43234, AU43235, AU43237, AU43240, AU49254, AU49256, AU49259, AV01003, AV01004, AV01005, AV01007, AV01009, AV01010, AV01011, AV01012, AV01013, AV01014, AV01016, AV06018, AV06020, AV06023, AV06026, AV06028, AV06030, AV10033, AV10034, AV10035, AV10036, AV10038, AV10039, AV10042, AV10044, AV10045, AV12047, AV12049, AV12054, AV12055, AV12056, AV14057, AV14059, AV14061, AV14062, AV14063, AV16072, AV16073, AV16075, AV16077, AV16078, AV16080, AV16082, AV16083, AV16086, AV16088, AV16089, AV16090, AV16091, AV21092, AV21093, AV21095, AV21096, AV22097, AV22100, AV22102, AV22105, AV22106, AV23107, AV23109, AV23110, AV23111, AV23112, AV23113, AV23116, AV27117, AV27118, AV27120, AV27121, AV36127, AV36128, AV36129, AV36130, AV39131, AV39132, AV39133, AV39134, AV39135, AV40136, AV40137, AV40139, AV40140, AV40143, AV40145, AV42147, AV42149, AV42150, AV42151, AV42152, AV42153, AV42154, AV42155, AV44156, AV44157, AV44158, AV44160, AV44161, AV44162, AV44163, AV44164, AV47167, AV47168, AV47169, AV47170, AV48171, AV48173, AV48174, AV48176, AV48177, AV48178, AV48179, AV49181, AV49183, AV49184, AV49185, AV49187, AV49188, AV49189, AV51191, AV51192, AW01001, AW01002, AW01007, AW01008, AW01010, AW01011, AW01012, AW01013, AW01014, AW01015, AW06016, AW06018, AW06019, AW06020, AW06021, AW06023, AW06024, AW06025, AW06026, AW06027, AW06028, AW06029, AW09031, AW09032, AW09033, AW09034, AW09035, AW10036, AW10037, AW10038, AW10039, AW10040, AW10041, AW10044, AW10045, AW14046, AW14048, AW14049, AW14051, AW16052, AW16053, AW16054, AW16056, AW16057, AW16058, AW16059, AW16060, AW16061, AW20062, AW20063, AW20064, AW20065, AW20066, AW20067, AW20068, AW20069, AW20070, AW20071, AW23072, AW23073, AW23074, AW23075, AW23076, AW23077, AW23078, AW23079, AW28087, AW28088, AW28089, AW28090, AW28091, AW28094, AW28095, AW28096, AW28097, AW28098, AW31099, AW31100, AW31102, AW31103, AW31104, AW31105, AW31106, AW31107, AW31108, AW31109, AW31110, AW35111, AW35112, AW35113, AW35114, AW35115, AW35116, AW35117, AW35120, AW35121, AW35122, AW35123, AW35124, AW35125, AW35126, AW35127, AW35128, AW42129, AW42131, AW42132, AW42133, AW45134, AW45135, AW45136, AW45137, AW45138, AW45139, AW45140, AW45141, AW45143, AW45144, AW45145, AW45146, AW45147, AW47148, AW47149, AW47151, AW47152, AW47153, AW47154, AW47155, AW47156, AW47157, AW49159, AW49160, AW49161, AW49162, AW49163, AW49164, AW49165, AY01001, AY01002, AY01004, AY01005, AY01006, AY01007, AY01008, AY01009, AY01010, AY04011, AY04012, AY04013, AY04015, AY04016, AY04017, AY04018, AY04019, AY04020, AY08021, AY08022, AY08023, AY08024, AY08025, AY08026, AY08027, AY08028, AY08029, AY08032, AY08033, AY08034, AY10036, AY10037, AY10038, AY10039, AY10040, AY10041, AY10042, AY10044, AY10045, AY10046, AY10047, AY10048, AY10049, AY10051, AY10053, AY10054, AY10055, AY17056, AY17057, AY17058, AY17059, AY17060, AY17061, AY17062, AY17063, AY17064, AY17065, AY19066, AY19067, AY19069, AY19070, AY19071, AY19072, AY19073, AY19074, AY19075, AY23076, AY23077, AY23078, AY23079, AY23080, AY23081, AY30085, AY30086, AY30087, AY30088, AY30090, AY32092, AY32093, AY32094, AY32095, AY32096, AY32097, AY32098, AY32099, AY32100, AY32101, AY32102, AY32103, AY32104, AY32105, AY32106, AY35107, AY35108, AY35109, AY35110, AY35111, AY35112, AY35113, AY35114, AY35115, AY35116, AY35117, AY35118, AY35119, AY35120, AY35121, AY40123, AY40124, AY40125, AY40126, AY41127, AY41128, AY41129, AY41130, AY41132, AY41133, AY41134, AY41140, AY41141, AY51142, AY51143, AY51144, AY51145, AY51149, AY52150, AY52151, AY52152, AY52153, AY52154, AY52155, AY52156, AY52157, AY52158, AY52159, AZ02001, AZ02002, AZ02004, AZ02006, AZ02007, AZ02008, AZ02009, AZ02010, AZ06011, AZ06012, AZ06013, AZ06014, AZ06015, AZ06016, AZ06017, AZ06018, AZ06020, AZ06021, AZ06022, AZ08023, AZ08024, AZ08025, AZ08026, AZ08027, AZ08028, AZ08029, AZ08030, AZ08031, AZ12033, AZ12034, AZ12035, AZ12038, AZ12040, AZ12042, AZ12043, AZ12044, AZ17045, AZ17046, AZ17048, AZ17050, AZ17051, AZ17053, AZ18054, AZ18055, AZ18056, AZ18057, AZ19058, AZ19059, AZ19060, AZ19061, AZ20062, AZ20063, AZ20064, AZ20065, AZ20066, AZ20067, AZ20068, AZ20069, AZ20070, AZ22071, AZ22072, AZ22073, AZ22074, AZ22075, AZ23077, AZ23078, AZ23079, AZ23080, AZ24081, AZ24082, AZ24083, AZ24084, AZ24085, AZ25087, AZ25088, AZ25089, AZ25090, AZ28091, AZ28092, AZ28093, AZ28094, AZ28095, AZ28096, AZ28097, AZ28098, AZ28099, AZ28100, AZ34101, AZ34102, AZ34103, AZ34104, AZ34105, AZ34106, AZ34107, AZ34108, AZ34109, AZ34110, AZ34111, AZ34112, AZ34113, AZ34114, AZ34115, AZ39116, AZ39118, AZ39119, AZ39120, AZ39121, AZ39122, AZ39123, AZ39124, AZ39125, AZ41127, AZ41131, AZ41132, AZ41133, AZ41134, AZ41135, AZ43136, AZ43137, AZ43138, AZ43139, AZ43140, AZ43141, AZ43143, AZ46146, AZ46147, AZ46148, AZ46149, AZ46150, AZ46151, AZ46152, AZ46156, AZ46157, AZ46158, AZ46159, AZ48160, AZ48162, AZ48163, AZ48164, AZ48166, AZ48167, AZ49168, AZ49169, AZ49170, AZ49172, AZ49173, AZ49174, AZ49175, AZ49176, BA03001, BA03002, BA03003, BA03004, BA03005, BA03006, BA03007, BA03008, BA03009, BA03010, BA07011, BA07012, BA07013, BA07014, BA07015, BA07016, BA07017, BA07018, BA07019, BA07020, BA08021, BA08022, BA08023, BA08024, BA08025, BA10026, BA10027, BA10028, BA10029, BA10030, BA10031, BA13032, BA13034, BA13035, BA13036, BA13037, BA13038, BA13039, BA13040, BA13041, BA15042, BA15045, BA16047, BA16048, BA16049, BA16050, BA17052, BA17053, BA17054, BA18055, BA18056, BA18057, BA19060, BA19061, BA19062, BA21064, BA22067, BA22068, BA22069, BA22070, BA23071, BA23072, BA23073, BA24075, BA24076, BA24077, BA24078, BA25080, BA25081, BA25082, BA25083, BA25084, BA25085, BA27086, BA27087, BA27088, BA27089, BA27090, BA29091, BA29092, BA29093, BA29094, BA29095, BA31096, BA31097, BA31098, BA31099, BA31100, BA35101, BA35103, BA35104, BA35105, BA36106, BA36107, BA37109, BA37110, BA37111, BA37112, BA38113, BA38114, BA38115, BA40116, BA40117, BA40118, BA41119, BA41120, BA41121, BA41122, BA43124, BA43126, BA46127, BA46128, BA46129, BA46130, BA48131, BA48132, BA48133, BA49134, BA49135, BA49136, BA49137, BA50138, BA50139, BA50140, BA50141, BA50142, BB03001, BB03002, BB03003, BB03004, BB04005, BB04006, BB05007, BB05008, BB05009, BB06010, BB09011, BB09012, BB09013, BB09014, BB09015, BB16016, BB16017, BB16018, BB16019, BB16021, BB16022, BB16023, BB23024, BB23025, BB23026, BB23027, BB23028, BB27029, BB27030, BB27031, BB27032, BB27033, BB27034, BB47045, BB47046, BB47047, BB47048, BB47049, BB47050, BB47051, BB47052, BB47053, BB47054, BC01001, BC01002, BC01003, BC01004, BC04006, BC04007, BC04008, BC10016, BC10017, BC10018, BC10019, BC14021, BC14022, BC14023, BC14024, BC21026, BC21027, BC29030, BC29031, BC29032, BC29035, BC30036, BC30039. Catalog Number B66705. Serial Numbers: AZ30002, AZ30003, AZ30004, AZ39005, BA14002. Catalog Number C11477. Serial Numbers: BB11109, BB11110, BB12111, BB12117, BB12118, BB12119, BB12120, BB12121, BB12122, BB12123, BB18124, BB18125, BB18126, BB18127, BB18128, BB18129, BB18130, BB18133, BB23134, BB23135, BB23136, BB31145, BB31146, BB31147, BB31148, BB31149, BB31150, BB36157, BB36158, BB36159, BB36160, BB36161, BB36162, BB43163, BB43164, BB43165, BB43166, BB43167, BB43168, BB47169, BB47170, BB47171, BB47172, BB47173, BB47174, BB47176, BB47177, BC01101, BC01102, BC01103, BC01104, BC01105, BC04106, BC04108, BC04109, BC04110, BC04111, BC04112, BC04113, BC04114, BC04115, BC09117, BC09118, BC09119, BC09120, BC09121, BC09122, BC09123, BC09124, BC09125, BC09126, BC09127, BC09128, BC14131, BC14132, BC14133, BC14134, BC14135, BC14136, BC14137, BC14138, BC14140, BC14141, BC14142, BC14143, BC14144, BC14145, BC19146, BC19147, BC19148, BC19149, BC19150, BC19151, BC19152, BC19153, BC19154, BC19155, BC19156, BC19157, BC21158, BC21159, BC21160, BC21161, BC21162, BC21163, BC21164, BC21166, BC21169, BC21171, BC24172, BC24173, BC24174, BC24175, BC24176, BC24178, BC24180, BC24181, BC30182, BC30183, BC30184, BC30185, BC30186, BC30187, BC30188, BC30189, BC30190, BC30191, BC33192, BC33193, BC33194, BC33195, BC33196, BC33197, BC33198, BC33200, BC33201, BC33202, BC33203, BC33204, BC33206, BC36207, BC36208, BC36209, BC36210, BC36211, BC36212, BC36213, BC36214, BC36215.

Distribution

Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Quantity

1,056 (418 US, and 638 OUS)