FDA Enforcement Class I Terminated

Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters. Labeling states manufactured by: Bunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use.

Recall: Z-0514-2013 · Reported December 26, 2012

Enforcement

Recall Number
Z-0514-2013
Event ID
63710
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Bunnell, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Press Release
Report Date
December 26, 2012
Initiation Date
November 19, 2012
Classification Date
December 14, 2012
Termination Date
February 21, 2014
Address
436 Lawndale Dr, Salt Lake City, UT, 84115-2917, United States

Description

Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters. Labeling states manufactured by: Bunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use.

Reason

Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.

Code Info

Patient Circuit (catalog # 902) lot # on each individual patient circuit label: 12C092, 12C102, 12C115, 12C125, 12C136, 12D159, 12D172, 12D189, 12E204, 12E211, 12E224, 12F241, 12F254, 12F271, 12G279, 12G290, 12G307, 12G321, 12H330, 12H349, 12I362, 12I371, 12I397, 12J413, 12J430, 12J448, 12K457, 12K471 Patient Circuit Kit (catalog # 397) lot # on the label on outside of box: 12C099, 12C114, 12C124, 12C135, 12D152, 12D170, 12D188, 12E203, 12E210, 12E223, 12F238, 12F255, 12F274, 12G209, 12G280, 12H334, 12H341, 12I356, 12I369, 12I392, 12J411, 12J425, 12J447, 12K467, 12K479

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, Malaysia, and Uruguay (animal study only).

Quantity

5743 patient circuits distributed in commerce