FDA Enforcement
Class II
Terminated
Total Temporomandibular Joint Replacement System- 45mm Rt Stand Off TI Mand Reconstruction of the temporomandibular joint
Recall: Z-0512-2015
·
Reported December 17, 2014
Enforcement
- Recall Number
- Z-0512-2015
- Event ID
- 69644
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomet Microfixation, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 17, 2014
- Initiation Date
- October 17, 2014
- Classification Date
- December 8, 2014
- Termination Date
- January 4, 2017
- Address
- 1520 Tradeport Dr, N/A, Jacksonville, FL, 32218-2480, United States
Description
Total Temporomandibular Joint Replacement System- 45mm Rt Stand Off TI Mand Reconstruction of the temporomandibular joint
Reason
Laser etching on the parts is wider and deeper than the conditions previously validated.
Code Info
Model 24-6645TI, Lot #s: 472580A, 472590B, 472580B, 472590A, 526030A, 526020A, 526020B, 526030B,
Distribution
Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.
Quantity
42