FDA Enforcement Class II Terminated

Total Temporomandibular Joint Replacement System- 45mm Lft Stand Off Ti Mand Reconstruction of the temporomandibular joint

Recall: Z-0510-2015 · Reported December 17, 2014

Enforcement

Recall Number
Z-0510-2015
Event ID
69644
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomet Microfixation, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 17, 2014
Initiation Date
October 17, 2014
Classification Date
December 8, 2014
Termination Date
January 4, 2017
Address
1520 Tradeport Dr, N/A, Jacksonville, FL, 32218-2480, United States

Description

Total Temporomandibular Joint Replacement System- 45mm Lft Stand Off Ti Mand Reconstruction of the temporomandibular joint

Reason

Laser etching on the parts is wider and deeper than the conditions previously validated.

Code Info

Models 24-6646TI, Lot #s: 526000A, 526000B, 526010B, 526010A, 535390A, 535390B, 535390C, 535390D, 538720A, 538720B, 538720C, 538720D

Distribution

Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.

Quantity

42