The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are indicated for cutting, grasping, dissecting, and coagulation of tissue in endoscopic and laparoscopic surgical procedures.
Enforcement
- Recall Number
- Z-0507-2018
- Event ID
- 78754
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Microline Surgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 14, 2018
- Initiation Date
- July 13, 2017
- Classification Date
- February 2, 2018
- Termination Date
- February 6, 2020
- Address
- 800 Cummings Ctr Ste 166T, N/A, Beverly, MA, 01915-6171, United States
Description
The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are indicated for cutting, grasping, dissecting, and coagulation of tissue in endoscopic and laparoscopic surgical procedures.
The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off during a surgical procedure and if cautery is needed, there is a potential for burn due to insulation failure.
All lots of product codes: 3201,3221,3231,3241,3251,3261,3271,3281,3291,3301,3311,3321,3331,3341, 3351,3361,3371,3381,3401,3411,3421,3431 UDI's for listed lots of product codes: 3201-UDI 00811099011644, 3221-UDI 00811099011651, 3231-UDI 00811099011668, 3241-UDI 00811099011675, 3251-UDI 00811099011682, 3261-UDI 00811099011699, 3271-UDI 00811099011705, 3281-UDI 00811099011712, 3291-UDI 00811099011729, 3301-UDI 00811099011859, 3311-UDI 00811099011736, 3321-UDI 00811099011743, 3331-UDI 00811099011767, 3341-UDI 00811099011774, 3351-UDI 00811099011781, 3371-UDI 00811099011804, 3381-UDI 00811099011811, 3401-UDI 00811099011828, 3411-UDI 00811099011835, 3421-UDI 00811099011842, 3431-UDI 00811099011644
Worldwide Distribution-US (Nationwide) and countries of: Australia, Brazil, Canada, Chile, China, Colombia, El Salvador, Hong Kong, India, Indonesia, Malaysia, Mexico, Panama, Peru, Argentina, South Korea, Sri Lanka, Taiwan, Thailand, Uruguay & Vietnam.
61,769 units