FDA Enforcement Class II Ongoing

Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.

Recall: Z-0506-2025 · Reported November 27, 2024

Enforcement

Recall Number
Z-0506-2025
Event ID
95601
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Zyno Medical LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 27, 2024
Initiation Date
October 18, 2024
Classification Date
November 21, 2024
Address
177 Pine St, N/A, Natick, MA, 01760-1331, United States

Description

Z-800 Infusion System. Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF. The Z-800 Infusion system is intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified healthcare professional.

Reason

Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.

Code Info

Model/Catalog Number: Z-800, Z-800F, Z-800W, Z-800WF Lot Code: Lots: 201016280, 200515888, 200406416, 230920212, 200413400, 20040660, 200716968, 201016720, 20071632, 201211740, 201208256, 210319640, 210709900, 201211360, 2101221000, 210913560, 210913200, 231008352, 220205400, 220204548, 2211141000, 230206980, 230620800, 230420548, 230420136, 230525452, 230620200.

Distribution

This product is distributed only in the United States.

Quantity

34,994 units