FDA Enforcement Class II Terminated

Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 100 Test 02816138 10309045, 10309044 028 Intended Use For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure.

Recall: Z-0506-2013 · Reported December 19, 2012

Enforcement

Recall Number
Z-0506-2013
Event ID
63682
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 19, 2012
Initiation Date
November 14, 2012
Classification Date
December 11, 2012
Termination Date
September 25, 2015
Address
333 Coney St, N/A, East Walpole, MA, 02032-1516, United States

Description

Siemens Healthcare BNP (8-type Natriuretic Peptide) 500 Test kit 02816634 100 Test 02816138 10309045, 10309044 028 Intended Use For in vitro diagnostic use in the quantitative determination of B-type Natriuretic Peptide (BNP) in human plasma using the ADVIA Centaur and ADVIA Centaur XP systems. This assay is indicated for the measurement of plasma BNP as an aid in the diagnosis and assessment of the severity of heart failure. In patients with acute coronary syndromes (ACS), this test, in conjunction with other known risk factors, can also be used to predict survival as well as to predict the likelihood of future heart failure.

Reason

Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Code Info

All lot codes

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to the following countries: Argentina Australia Austria Azerbaijan Belgium Bosnia Herzeg. Brazil Canada Chile COLOMBIA Croatia Czech Republic Denmark Egypt Estonia Finland France Germany Greece Hong Kong Hungary Iceland India Indonesia Iran Ireland Italy Japan Kazakhstan Latvia Lithuania Malaysia Mexico Netherlands New Caledonia Norway Oman P.R. China Philippines Poland Portugal Qatar Rep. of Yemen Republic Korea Reunion Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. United Kingdom Uruguay Vietnam

Quantity

100 test=24291 and 500 test= 9679