Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on victims of sudden cardiac arrest (SCA.)
Enforcement
- Recall Number
- Z-0505-2025
- Event ID
- 95594
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Defibtech, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 4, 2024
- Initiation Date
- October 14, 2024
- Classification Date
- November 21, 2024
- Address
- 741 Boston Post Rd, Suite 201, Guilford, CT, 06437-2714, United States
Description
Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on victims of sudden cardiac arrest (SCA.)
Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
UDI-DI: 00815098020454, DDP-2002, pediatric defibrillation pads, single pair; 10815098020451, DDP-2002, pediatric defibrillation pads, single pair configuration; 40815098020452, DDP-2002, pediatric defibrillation pads, 20 pairs
Ireland
22,294 total