FDA Enforcement Class II Ongoing

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Recall: Z-0501-2025 · Reported November 27, 2024

Enforcement

Recall Number
Z-0501-2025
Event ID
95581
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
OrganOx Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2024
Initiation Date
October 15, 2024
Classification Date
November 21, 2024
Address
Magdalen Centre, 1 Robert Robinson Avenue, Oxford, N/A, N/A, United Kingdom

Description

OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.

Reason

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Code Info

UDI-DI: 05060462240029 Serial numbers 511, 512, 514, 515, 516, 517, 518, 519, 520, 522, 523, 524, 525, 527, 528, 529, 530, 531, 533, 535, 536, 537

Distribution

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

Quantity

20 units