FDA Enforcement Class II Ongoing

Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Picture Archiving and Communication System) solutions for hospital and related institutions/sites.

Recall: Z-0497-2025 · Reported November 27, 2024

Enforcement

Recall Number
Z-0497-2025
Event ID
95604
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips Medical Systems Technologies Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2024
Initiation Date
October 10, 2024
Classification Date
November 20, 2024
Address
Matam Building 8/2, Advanced Technology Center, Haifa, N/A, N/A, Israel

Description

Vue PACS Diagnostic Viewer with Vue PACS client 12.1.5 to 12.2.8.454 software Model Number: 1149418 Intended use is to provide completely scalable local and wide area PACS (Picture Archiving and Communication System) solutions for hospital and related institutions/sites.

Reason

A software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to incorrect ischemic map and table values.

Code Info

UDI-DI: (01)00884838100350(11)200221(10)12.2.5; (01)00884838100350(11)200812(10)12.2.6.0; (01)00884838100350(11)201207(10)12.2.6.100; (01)00884838100350(11)210303(10)12.2.6.200; (01)00884838100350(11)210518(10)12.2.6.201; (01)00884838100350(11)210729(10)12.2.6.300; (01)00884838100350(11)210502(10)12.2.8.0; (01)00884838100350(11)220112(10)12.2.8.100; (01)00884838100350(11)221213(10)12.2.8.200; (01)00884838100350(11)230301(10)12.2.8.210; (01)00884838100350(11)230320(10)12.2.8.300; (01)00884838100350(11)230608(10)12.2.8.310; (01)00884838100350(11)230718(10)12.2.8.400; (01)00884838100350(11)230810(10)12.2.8.401; (01)00884838100350(11)230824(10)12.2.8.317; (01)00884838100350(11)231019(10)12.2.8.412; (01)00884838100350(11)231108(10)12.2.8.403; (01)00884838100350(11)231129(10)12.2.8.413; (01)00884838100350(11)231214(10)12.2.8.414; (01)00884838100350(11)231204(10)12.2.8.420; (01)00884838100350(11)240101(10)12.2.8.415; (01)00884838100350(11)240116(10)12.2.8.421; (01)00884838100350(11)240204(10)12.2.8.422; (01)00884838100350(11)240226(10)12.2.8.424; (01)00884838100350(11)240311(10)12.2.8.426; (01)00884838100350(11)240407(10)12.2.8.440; (01)00884838100350(11)240425(10)12.2.8.450; (01)00884838100350(11)240613(10)12.2.8.451; (01)00884838100350(11)240715(10)12.2.8.452; (01)00884838100350(11)240715(10)12.2.8.454; (01)00884838100350(11)230926(10)12.2.8.410

Distribution

US Nationwide and Worldwide Distribution Foreign: Country Argentina Australia Austria Belgium Brazil Chile China Colombia Denmark Finland France Germany Gibraltar Greece Guernsey Holy See (Vatican City State) Hong Kong Hungary Indonesia Ireland Israel Italy Japan Jersey Jordan Lebanon Malaysia Mexico Netherlands New Zealand Norway Oman Panama Philippines Poland Portugal Romania Saudi Arabia Serbia Slovenia Spain Sweden Switzerland Taiwan Thailand United Kingdom

Quantity

2439 systems