FDA Enforcement
Class II
Ongoing
STRADIS HEALTHCARE, Implants, Item No.41015NSI, dental tray
Recall: Z-0493-2023
·
Reported December 21, 2022
Enforcement
- Recall Number
- Z-0493-2023
- Event ID
- 91124
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stradis Medical, LLC dba Stradis Healthcare
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 21, 2022
- Initiation Date
- November 2, 2022
- Classification Date
- December 14, 2022
- Address
- 3025 Northwoods Pkwy, N/A, Peachtree Corners, GA, 30071-1524, United States
Description
STRADIS HEALTHCARE, Implants, Item No.41015NSI, dental tray
Reason
Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.
Code Info
UDI/DI (case) M75241015NSI1, UDI/DI (kit)M75241015NSI0, Serial/Lot Numbers: 22230491043
Distribution
US nationwide distribution, and Canada.
Quantity
30 units