FDA Enforcement
Class II
Terminated
10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Recall: Z-0493-2018
·
Reported February 7, 2018
Enforcement
- Recall Number
- Z-0493-2018
- Event ID
- 78692
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Horiba Instruments, Inc dba Horiba Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 7, 2018
- Initiation Date
- November 21, 2017
- Classification Date
- January 31, 2018
- Termination Date
- November 4, 2019
- Address
- 34 Bunsen, Irvine, CA, 92618-4210, United States
Description
10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Reason
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Code Info
DATE - MVT 25.08.2017
Distribution
Nationwide
Quantity
10 bags, 100 units each