FDA Enforcement Class II Terminated

10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400

Recall: Z-0493-2018 · Reported February 7, 2018

Enforcement

Recall Number
Z-0493-2018
Event ID
78692
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Horiba Instruments, Inc dba Horiba Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 7, 2018
Initiation Date
November 21, 2017
Classification Date
January 31, 2018
Termination Date
November 4, 2019
Address
34 Bunsen, Irvine, CA, 92618-4210, United States

Description

10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400

Reason

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Code Info

DATE - MVT 25.08.2017

Distribution

Nationwide

Quantity

10 bags, 100 units each