FDA Enforcement Class II Terminated

BD Vacutainer Specimen Collection Cups, Cat. No. 364975

Recall: Z-0492-2019 · Reported November 28, 2018

Enforcement

Recall Number
Z-0492-2019
Event ID
81148
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 28, 2018
Initiation Date
August 30, 2018
Classification Date
November 21, 2018
Termination Date
October 21, 2020
Address
1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States

Description

BD Vacutainer Specimen Collection Cups, Cat. No. 364975

Reason

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Code Info

All lots within expiry

Distribution

Illinois

Quantity

Total: 29,982,150 BD cups (US); 97,800,900 BD cups (OUS)