FDA Enforcement
Class II
Terminated
Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Recall: Z-0492-2018
·
Reported February 7, 2018
Enforcement
- Recall Number
- Z-0492-2018
- Event ID
- 78776
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Tosoh Bioscience Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 7, 2018
- Initiation Date
- June 11, 2017
- Classification Date
- January 30, 2018
- Termination Date
- December 31, 2019
- Address
- 3600 Gantz Rd, Grove City, OH, 43123-1895, United States
Description
Tosoh¿Sorter 9tray, Product Code 022931and Tosoh¿Sorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.
Reason
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Code Info
AIA-900 19tray Sorter with S/N 10000109 to S/N 80136909. AIA-900 9tray Sorter with S/N 10000109 to S/N 10707402.
Distribution
Worldwide Distribution - U.S. Nationwide - Latin America
Quantity
565 total devices U.S.