FDA Enforcement
Class II
Terminated
Geistlich Bio-Oss Pen 0.25-1mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20116
Recall: Z-0491-2021
·
Reported December 2, 2020
Enforcement
- Recall Number
- Z-0491-2021
- Event ID
- 86628
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Geistlich Pharma North America, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 2, 2020
- Initiation Date
- October 19, 2020
- Classification Date
- November 25, 2020
- Termination Date
- July 13, 2021
- Address
- 202 Carnegie Ctr Ste 103, N/A, Princeton, NJ, 08540-6239, United States
Description
Geistlich Bio-Oss Pen 0.25-1mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20116
Reason
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
Code Info
Lots: 81901649, 82000457, 82000633
Distribution
Worldwide distribution - US Nationwide distribution and the country of Japan.
Quantity
2097 units