STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
Enforcement
- Recall Number
- Z-0490-2024
- Event ID
- 93300
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stanbio Laboratory, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 13, 2023
- Initiation Date
- October 2, 2023
- Classification Date
- December 6, 2023
- Address
- 1261 N Main St, Boerne, TX, 78006-3014, United States
Description
STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.
UDI-DI: 00657498000090, Lot/Expiration: 203044/2024-01-31, 209644/2024-06-30
Worldwide - US Nationwide distribution including in the states of AZ, RI, MO, UT, AK, LA, TX, OH, GA, MN, NY, CA, NJ, IL, WA, KS, MI, NC, PA, MA, CO, KY, NM, FL, MD, VA, IN, AL, SD, MS, WV, WI, ID, MT, WY and the countries of Canada and Singapore.