FDA Enforcement Class II Ongoing

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

Recall: Z-0489-2025 · Reported November 27, 2024

Enforcement

Recall Number
Z-0489-2025
Event ID
95609
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2024
Initiation Date
October 17, 2024
Classification Date
November 19, 2024
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

Reason

Abnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)

Code Info

UDI/DI 04015630929979, Lot Number 747029, exp. 30-SEPT-2024

Distribution

US: NJ, TX, NY, FL, IN, TX, OK, SC, IA,

Quantity

458 units