FDA Enforcement
Class II
Ongoing
Sedecal NOVA FA DR System. for diagnostic radiography.
Recall: Z-0488-2019
·
Reported November 28, 2018
Enforcement
- Recall Number
- Z-0488-2019
- Event ID
- 81377
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Sedecal S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 28, 2018
- Initiation Date
- July 18, 2018
- Classification Date
- November 20, 2018
- Address
- Pol. Industrial Rio De Janiero, Calle Pelaya 9-13, Algete, N/A, N/A, Spain
Description
Sedecal NOVA FA DR System. for diagnostic radiography.
Reason
A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
Code Info
Serial Numbers 07360816 through 17291024
Distribution
US and Canada
Quantity
145 units