FDA Enforcement Class II Ongoing

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;

Recall: Z-0486-2026 · Reported November 19, 2025

Enforcement

Recall Number
Z-0486-2026
Event ID
97792
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 19, 2025
Initiation Date
September 25, 2025
Classification Date
November 7, 2025
Address
60 Middletown Ave, North Haven, CT, 06473-3908, United States

Description

Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;

Reason

A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.

Code Info

CFN: BOX06170V4; UDI-DI: 00763000988975; Lot numbers: 0231961695;

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.

Quantity

2 units (OUS only)