FDA Enforcement Class II Ongoing

Ascom Telligence Nurse Call System

Recall: Z-0486-2019 · Reported November 28, 2018

Enforcement

Recall Number
Z-0486-2019
Event ID
81546
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ascom (US) Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
November 28, 2018
Initiation Date
November 2, 2018
Classification Date
November 20, 2018
Address
9024 Town Center Parkway, N/A, Lakewood Ranch, FL, 34202-4159, United States

Description

Ascom Telligence Nurse Call System

Reason

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Code Info

Version 5.0 (Build 5.0.4)

Distribution

AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI

Quantity

57 units