FDA Enforcement
Class II
Ongoing
Ascom Telligence Nurse Call System
Recall: Z-0486-2019
·
Reported November 28, 2018
Enforcement
- Recall Number
- Z-0486-2019
- Event ID
- 81546
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ascom (US) Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- November 28, 2018
- Initiation Date
- November 2, 2018
- Classification Date
- November 20, 2018
- Address
- 9024 Town Center Parkway, N/A, Lakewood Ranch, FL, 34202-4159, United States
Description
Ascom Telligence Nurse Call System
Reason
There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.
Code Info
Version 5.0 (Build 5.0.4)
Distribution
AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI
Quantity
57 units