FDA Enforcement Class II Ongoing

PUMP MCS1705PU HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT US, Model Number MCS1705PU

Recall: Z-0483-2024 · Reported December 13, 2023

Enforcement

Recall Number
Z-0483-2024
Event ID
93405
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Heartware, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
December 13, 2023
Initiation Date
October 31, 2023
Classification Date
December 1, 2023
Address
14400 Nw 60th Ave, Miami Lakes, FL, 33014-2807, United States

Description

PUMP MCS1705PU HVAD (HeartWare Ventricular Assist Device) IMPLANT KIT US, Model Number MCS1705PU

Reason

Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components

Code Info

GTIN 00888707005364, Serial Numbers: HW43366, HW42611, HW43316, HW43247, HW43248, HW42405, HW43276, HW43251, HW43313, HW43322, HW43257, HW43262, HW43258, HW43270, HW43197, HW43157, HW43253, HW42375, HW42400, HW42582, HW43125, HW42083, HW43188, HW43042, HW42610, HW42301, HW43062, HW43219, HW43187, HW42420, HW43054, HW43115, HW43053, HW43116, HW42297, HW43109, HW42267, HW43048, HW42591, HW42439, HW42205, HW43040, HW42438, HW42586, HW42082, HW42585, HW42261, HW42404, HW42573, HW42364, HW42070, HW42202, HW42081, HW42184, HW42179, HW42069, HW42403, HW42176, HW41765; GTIN 00888707005715, Serial Numbers: HW42612, HW42685, HW42614, HW42681, HW42615, HW42618; GTIN 00888707007139, Serial Numbers: HW42372, HW42691, HW42832, HW42713, HW43019, HW43027, HW43015, HW42847, HW42697, HW42963, HW42826, HW42929, HW42934, HW42975, HW42954, HW42845, HW42705, HW42846, HW42925, HW42977, HW42978, HW42839, HW42979, HW42980, HW42982, HW42984, HW42985, HW42986, HW42704, HW42930, HW42972, HW42853, HW42924, HW42935, HW42936, HW42956, HW42957, HW42958, HW42959, HW42945, HW42946, HW42950, HW42951, HW42706, HW42699, HW42701, HW42692, HW42693, HW42694, HW42695, HW42696; GTIN 00888707007160, Serial Numbers: HW43453; GTIN 00888707009041, Serial Numbers: HW43032, HW42484, HW42522, HW43489; GTIN 00888707009423, Serial Numbers: HW43068; GTIN 00888707009454, Serial Numbers: HW43096; GTIN 00888707009485, Serial Numbers: HW42688; GTIN 00888707009492, Serial Numbers: HW42926

Distribution

Worldwide

Quantity

125