FDA Enforcement Class II Ongoing

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Recall: Z-0482-2026 · Reported November 19, 2025

Enforcement

Recall Number
Z-0482-2026
Event ID
97733
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 19, 2025
Initiation Date
October 6, 2025
Classification Date
November 7, 2025
Address
3500 Corporate Pkwy, Center Valley, PA, 18034-8229, United States

Description

Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of diminutive polyps, sessile polyps, pedunculated polyps, and tissue from within the GI tract.

Reason

Sterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.

Code Info

Model: SD-400U-10; Lot Numbers: 27V, 28V, 29V, 2XV, 2YV, 2ZV, 31V, 32V, 33V, 34V, 35V, 36V, 37V, 38V, 39V, 3XV, 3YV, 3ZV, 41V, 42V, 43V, 44V, 45V, 46V, 47V, 48V, 49V, 4XV, 4YV, 51V, 52V, 53V, 54V, 55V, 56V, 57V UDI: 04953170408243

Distribution

Worldwide distribution - US Nationwide and the countries of GU, and MP.

Quantity