FDA Enforcement Class II Terminated

Insulia Diabetes Management Companion

Recall: Z-0478-2019 · Reported November 28, 2018

Enforcement

Recall Number
Z-0478-2019
Event ID
81508
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Voluntis
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
November 28, 2018
Initiation Date
October 8, 2018
Classification Date
November 16, 2018
Termination Date
March 12, 2019
Address
22 Quai Gallieni, N/A, Suresnes, N/A, N/A, France

Description

Insulia Diabetes Management Companion

Reason

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

Code Info

Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40

Distribution

Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).

Quantity

14