FDA Enforcement Class II Ongoing

OCT-Camera 211 01 A1, Running software iOCT-Control 3 v1.4

Recall: Z-0477-2022 · Reported January 19, 2022

Enforcement

Recall Number
Z-0477-2022
Event ID
89176
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Haag-Streit USA Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 19, 2022
Initiation Date
November 3, 2021
Classification Date
January 10, 2022
Address
3535 Kings Mills Rd, N/A, Mason, OH, 45040-2303, United States

Description

OCT-Camera 211 01 A1, Running software iOCT-Control 3 v1.4

Reason

Malfunction of the automatic laser beam shut-off, the OCT unit might not recognize whether the laser beam is safely switched off

Code Info

Serial no. 0005, 0023

Distribution

US Nationwide distribution in the states of OH, IN.

Quantity

2 devices quarantined in US