Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
Enforcement
- Recall Number
- Z-0476-2023
- Event ID
- 91200
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Leica Microsystems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 21, 2022
- Initiation Date
- November 21, 2022
- Classification Date
- December 9, 2022
- Address
- 10 Parkway North blvd ste 300, N/A, deerfield, IL, 60089-6622, United States
Description
Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system
There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.
***EXPANSION 09/08/2023 UPDATE a. Premium Part Number 10448950; Serial Numbers: 160222001, 280622001, 50422001, 50422002, 91221001, 180622001, 170622001, 61221001, 130922001, 120623001, 150623001, 81022001, 200623001, 51022001, 61022001. b. Standard Part Number 10448976; Serial Numbers: 120322001, 300122001, 40222002, 301221001, 40122001, 140522001, 151221002, 81221002, 100222002, 300122002, 110322001, 50322001, 90622001, 120422001, 90522001, 100522001, 250122001, 100222001, 171022001, 80922001, 70623001, 30922001, 91022001.
US Nationwide
38 units