BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Enforcement
- Recall Number
- Z-0473-2023
- Event ID
- 91201
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Becton Dickinson & Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 21, 2022
- Initiation Date
- November 21, 2022
- Classification Date
- December 9, 2022
- Address
- 7 Loveton Cir, N/A, Sparks, MD, 21152-9212, United States
Description
BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
UDI: (10)2206461(17)221125(30)0100 Lot Number: 2206461 Exp. Date: 20221125. Box numbers 0002 through 0097. Boxes correlate to timestamps on the plates of 11:34 through 13:51. All other box numbers and plate stamps are unaffected and perform as intended. (Timestamps can be found on the bottom of the plate. The box number can be located on the carton label).
Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.
9,600 EA