FDA Enforcement Class II Terminated

Vancomycin Screen Agar

Recall: Z-0473-2016 · Reported December 30, 2015

Enforcement

Recall Number
Z-0473-2016
Event ID
71883
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Co.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 30, 2015
Initiation Date
June 29, 2015
Classification Date
December 23, 2015
Termination Date
August 12, 2016
Address
BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD, 21152-0999, United States

Description

Vancomycin Screen Agar

Reason

The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.

Code Info

catalog number 222204. Lot numbers 5041659, 5064921, 5076659, 5083712, 5097504, 5105955, 5130907, 5139626, and 5147941

Distribution

Worldwide Distribution - US (nationwide) Distribution to the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI and WV., and to the countries of : Canada, Colombia, India, Mexico and Europe.

Quantity

60,840