FDA Enforcement Class II Ongoing

ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

Recall: Z-0472-2023 · Reported December 21, 2022

Enforcement

Recall Number
Z-0472-2023
Event ID
91182
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 21, 2022
Initiation Date
November 9, 2022
Classification Date
December 9, 2022
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

ARTISTE, ONCOR and PRIMUS with syngo RT Therapist with software version 4.3.1 MR3. Material Number: 8162815 Intended Use: Family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

Reason

When selecting the site fraction group for treatment, the User may select the wrong site for treatment, and potentially deliver the dose to wrong isocenter

Code Info

UDI: N/A S/N: 10736, 10737, 10822, 10850, 10886,10843, 10865,10025 Software version 4.3.1 MR3

Distribution

US Nationwide distribution in the states of CA, IA, LA, NE, NV, OH, SC.

Quantity

8 units US