FDA Enforcement Class II Terminated

OSS Reamer Sleeve REF132-472662 KNEE REAMER SLEEVE 13.0 thru 20.0 STAINLESS STEEL The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.

Recall: Z-0472-2013 · Reported December 12, 2012

Enforcement

Recall Number
Z-0472-2013
Event ID
63612
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
December 12, 2012
Initiation Date
November 1, 2012
Classification Date
December 4, 2012
Termination Date
April 29, 2013
Address
56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States

Description

OSS Reamer Sleeve REF132-472662 KNEE REAMER SLEEVE 13.0 thru 20.0 STAINLESS STEEL The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.

Reason

Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.

Code Info

Part Numbers 32-472662, 13.0 thru 20.0 Lot 154590, 156980, 519470, 519480, 542290, 628890, 755530

Distribution

Worldwide distribution: USA (nationwide) and countries of: Japan and Netherlands.

Quantity

36