FDA Enforcement Class III Ongoing

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474

Recall: Z-0471-2021 · Reported November 25, 2020

Enforcement

Recall Number
Z-0471-2021
Event ID
86669
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Ortho Clinical Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 25, 2020
Initiation Date
October 16, 2020
Classification Date
November 18, 2020
Address
1001 US Highway 202, N/A, Raritan, NJ, 08869-1424, United States

Description

VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474

Reason

Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures

Code Info

Lot Q7692 , EXP: 01-NOV-2021 UDI: 10758750004577

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Quantity

10788 total: US 4680; OUS 6108